Static forces of body weight and dynamic forces induced by walking are a challenge toward lasting fixation devices in the case of unstable peritrochanteric fractures. Experiences with the newly developed gamma nail internal fixation technique are promising. The fixation is stable, complication rates are low, and fractures of the gamma nail itself have not been reported before. From the Dutch experience, the authors describe the four known cases of implant failure. In cases in which the process of bony healing is disturbed, the dynamic stresses during ambulation become large enough to cause metal fatigue fractures. In these cases adjuvant and appropriate measures should be undertaken.
对于不稳定的转子周围骨折,体重产生的静力和行走引起的动力对持久固定装置构成了挑战。新开发的伽马钉内固定技术的经验很有前景。该固定稳定,并发症发生率低,且此前尚未有伽马钉本身骨折的报道。根据荷兰的经验,作者描述了4例已知的植入物失败病例。在骨愈合过程受到干扰的情况下,行走时的动态应力会变得足够大,从而导致金属疲劳骨折。在这些情况下,应采取辅助和适当的措施。