Suppr超能文献

用于测量步态时间参数的脚踏开关系统。

Footswitch system for measurement of the temporal parameters of gait.

作者信息

Hausdorff J M, Ladin Z, Wei J Y

机构信息

Department of Biomedical Engineering, Boston University, Massachusetts, USA.

出版信息

J Biomech. 1995 Mar;28(3):347-51. doi: 10.1016/0021-9290(94)00074-e.

Abstract

Gait analysis relies upon accurate measurement of initial and end foot contact times. These times act as a reference point for correlating all other gait data and as a mean of distinguishing normal and pathologic gait. We have developed a simple, inexpensive footswitch system that provides accurate estimates of the start and end of stance phase for sequential steps. The estimates of the beginning and end of stance phase do not require custom footwear, extensive calibration, or precise placement of the sensor within the shoe. The system is based on a commercially available transducer and can be readily reproduced for use in a laboratory setting for less than $50. We describe this system, as well as its validation. To assess the accuracy of this footswitch system, we compared footswitch based estimates of initial and end foot contact times with those obtained using a force platform as 10 people took 30 steps (10 each at slow, normal and fast walking rates) across a force platform. Both estimates coincided within +/- 10 ms (mean: 0 +/- 3 ms; N = 300) for the start of stance phase and within +/- 22 ms (mean: -1 +/- 8 ms; N = 300) for the end of stance phase. For stance duration, the differences ranged from -24 to 28 ms (mean: 1 +/- 10 ms; N = 300). In combination, these measures can be used to estimate stance duration to within 3% of force plate determined values for steps with stance durations ranging from 446 to 1594 ms. Estimates of swing and stride duration also are within 5% of force plate determined values. This system should therefore prove to be a useful tool for augmenting laboratory based investigations of gait.

摘要

步态分析依赖于对初始和终末足部接触时间的精确测量。这些时间作为关联所有其他步态数据的参考点,并作为区分正常和病理性步态的一种手段。我们开发了一种简单、廉价的脚踏开关系统,该系统能为连续的步态准确估计站立期的开始和结束。站立期开始和结束的估计不需要定制鞋具、广泛校准或传感器在鞋内的精确放置。该系统基于一种商用传感器,在实验室环境中易于复制,成本不到50美元。我们描述了这个系统及其验证过程。为了评估这个脚踏开关系统的准确性,我们让10个人以慢、正常和快三种步行速度在测力平台上各走30步(每种速度各走10步),将基于脚踏开关的初始和终末足部接触时间估计值与使用测力平台获得的估计值进行比较。对于站立期开始,两种估计值在±10毫秒内一致(平均值:0±3毫秒;N = 300),对于站立期结束,在±22毫秒内一致(平均值:-1±8毫秒;N = 300)。对于站立持续时间,差异范围为-24至28毫秒(平均值:1±10毫秒;N = 300)。综合起来,这些测量可用于将站立持续时间估计在测力平台确定值的3%以内,适用于站立持续时间为446至1594毫秒的步态。摆动期和步幅持续时间的估计值也在测力平台确定值的5%以内。因此,该系统应被证明是增强基于实验室的步态研究的有用工具。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验