Somer T, Milicić A, Lazetić A, Brankov M
Medicinski fakultet, Institut za hirurgiju, Novi Sad.
Med Pregl. 1994 Jan-Feb;47(1-2):59-61.
This is a case report on a patient with a multiple femur's fracture where a transfixed intramedullar nailing of Grosse-Kempf type has been done. "The nail dynamization" was not carried out before the beginning of vertical loading. Seven months after the full vertical loading had begun, at a sudden torsian movement of the extremity, the nail had broken. Analyzing the cause of the fixative material fracture, a conclusion has been made that, apart from the presence of adequate trauma, the nail fracture happened also due to the development of pseudoarthrosis of femur and the unfulfilled "nail dynamization".
这是一份关于一名股骨多处骨折患者的病例报告,该患者接受了Grosse-Kempf型髓内钉固定术。在开始垂直负重之前未进行“髓内钉动力化”操作。在完全开始垂直负重七个月后,肢体突然扭转时,髓内钉断裂。通过分析固定材料断裂的原因得出结论,除了存在足够的创伤外,髓内钉断裂还由于股骨假关节的形成以及“髓内钉动力化”未实施。