Abstract

People with heart problems have had their lives extended considerably with the development of the prosthetic heart valve. Great strides have been made in the development of the valves through the use of improved materials as well as efficient mechanical designs. However, since the valves operate continuously over a long period, structural failures can occur--even though they are relatively uncommon. Here the development of techniques to classify the valve either as having intact struts or as having a separated strut, commonly called single leg separation, is discussed. In this paper the signal processing techniques employed to extract the required signals/parameters are briefly reviewed and then it is shown how they can be used to simulate a synthetic heart valve database for eventual Monte Carlo testing. Next, the optimal classifier is developed under assumed conditions and its performance is compared to that of an adaptive-type classifier implemented with a probabilistic neural network. Finally, the adaptive classifier is applied to a data set and its performance is analyzed. Based on synthetic data it is shown that excellent performance of the classifiers can be achieved implying a potentially robust solution to this classification problem.