Temperature may be an appropriate sensor for chronotropically incompetent patients with postural syncope.

作者信息

Singer I, Ripley K, Johnson B, Stoddard M

机构信息

University of Louisville, Kentucky 40292.

出版信息

Pacing Clin Electrophysiol. 1994 Oct;17(10):1655-64. doi: 10.1111/j.1540-8159.1994.tb02360.x.

Abstract

Chronotropically incompetent patients benefit most from sensor driven rate response during exercise. Postural syncope may occur despite the chronotropic response because of the failure of currently available sensors to respond physiologically to postural changes. Seven chronotropically incompetent patients with postural syncope who had a dual chamber rate adaptive pacemaker (Circadia) that modulates heart rate in response to temperature change were studied with respect to: (1) response to exercise; and (2) head-up tilt (HUT). During exercise, continuous-wave Doppler of aortic velocities and two-dimensional echocardiographic derived measurements of left ventricular systolic function were used to assess cardiac function. Patients exercised longer (by an average of 168 sec) in the DDDR compared to the DDI mode (P = 0.013). Increase in exercise duration was due mostly to the sensor driven increase during DDDR pacing. During DDDR pacing, heart rate increased from 71 +/- 6 to 121 +/- 17 ppm compared to 70 +/- 1 to 103 +/- 21 ppm for the DDI pacing (P = 0.038). Stroke volume as assessed by Doppler derived stroke distance (SD) contributed more significantly to the cardiac output increase during exercise in the DDI mode (SD increased from 13.4 +/- 4 to 18 +/- 7 cm in DDI compared to 13 +/- 4 to 14 +/- 2 cm in DDDR mode), although these mechanisms were insufficient to fully compensate for failure of appropriate chronotropic response. In response to the HUT, right ventricular temperature increased from 36.78 degrees C +/- 0.29 degrees C to 36.89 degrees +/- 0.28 degrees C (P = 0.0002), and heart rate increased from 54 +/- 3 to 71 +/- 8 ppm (P = 0.0003) in the DDDR mode. No significant change in heart rate occurred in the DDI mode in response to the HUT. Strong positive correlation of temperature and heart rate was noted in all patients in response to HUT (P = 0.001, R2 = 0.755-0.976). We conclude that temperature sensor responds physiologically to exercise and HUT. Therefore, temperature sensing rate adaptive dual chamber pacing may be appropriate for chronotropically incompetent patients with posture related syncope.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验