Suppr超能文献

新型单通道VDD起搏系统的临床评估

Clinical evaluation of a new single pass lead VDD pacing system.

作者信息

Ovsyshcher I E, Katz A, Bondy C

机构信息

Division of Cardiology, Soroka Medical Center, Beer-Sheva, Israel.

出版信息

Pacing Clin Electrophysiol. 1994 Nov;17(11 Pt 2):1859-64. doi: 10.1111/j.1540-8159.1994.tb03763.x.

Abstract

UNLABELLED

Twenty-five patients with second- to third-degree AV block and normal sinus function (16 males, mean age 60 +/- 18; range 15-78 years) underwent implantation of VVD pacemakers (THERA VDD, Medtronic, Inc.) with a single pass (SP) lead.

RESULTS

During implantation the mean amplitude of the atrial (A) signal was 3.9 +/- 1.4 mV (range 2.0-7.8 mV). Stable, acceptable A-signals during implantation were usually observed in the mid- or lower part of the right atrium. The lead tip electrical parameters were not compromised in any patient in order to obtain an acceptable A-signal. To verify VDD device function, patients underwent pacing system analysis on the second day and again 1, 3, and 6 months after implantation. Acute and chronic electrical measurements in the ventricle were similar to those with regular steroid leads. During follow-up tests, stable atrial sensing (A > or = 0.7 mV) was found in all but one patient (in whom A was 0.25-0.5 mV and an intermittent loss of atrial sensing occurred). There was no difference between serial measurements of A-signal amplitudes on the second day or 1, 3, and 6 months after implantation: 1.9 +/- 1.3 mV, 1.5 +/- 0.6 mV, 1.3 +/- 0.8 mV, and 1.5 +/- 1.1 mV, respectively. The mean implantation time was 54.0 +/- 17 minutes and the mean fluoroscopy time was 3.2 +/- 1.3 minutes.

CONCLUSIONS

SP lead VDD pacing is reliable and easy to manage with dependable atrial sensing and ventricular pacing.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

未标注

25例二度至三度房室传导阻滞且窦性功能正常的患者(16例男性,平均年龄60±18岁;年龄范围15 - 78岁)接受了单极(SP)导线的VVD起搏器(美敦力公司的THERA VDD)植入术。

结果

植入过程中,心房(A)信号的平均幅度为3.9±1.4 mV(范围2.0 - 7.8 mV)。植入过程中稳定、可接受的A信号通常在右心房中部或下部观察到。为获得可接受的A信号,没有任何患者的导线尖端电参数受到损害。为验证VDD装置功能,患者在植入后第二天以及1、3和6个月再次接受起搏系统分析。心室的急性和慢性电测量结果与常规类固醇导线的测量结果相似。在随访测试中,除1例患者(其A信号为0.25 - 0.5 mV且发生间歇性心房感知丧失)外,所有患者均发现稳定的心房感知(A≥0.7 mV)。植入后第二天、1、3和6个月的A信号幅度系列测量之间无差异:分别为1.9±1.3 mV、1.5±0.6 mV、1.3±0.8 mV和1.5±1.1 mV。平均植入时间为54.0±17分钟,平均透视时间为3.2±1.3分钟。

结论

SP导线VDD起搏可靠且易于管理,具有可靠的心房感知和心室起搏功能。(摘要截断于250字)

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验