Suppr超能文献

重症监护病房中醋酸钙不动杆菌无硝亚种暴发,源头追溯至一个峰值流量计。

Acinetobacter calcoaceticus anitratus outbreak in the intensive care unit traced to a peak flow meter.

作者信息

Ahmed J, Brutus A, D'Amato R F, Glatt A E

机构信息

Department of Infection Control, Catholic Medical Center of Brooklyn and Queens, Inc., Jamaica, NY 11432.

出版信息

Am J Infect Control. 1994 Oct;22(5):319-21. doi: 10.1016/0196-6553(94)90020-5.

Abstract

BACKGROUND

A cluster of seven cases of Acinetobacter caleoaceticus anitratus in a community teaching hospital intensive care unit was discovered (the seventh case was located in a step-down unit next to an infected patient recently transferred from the intensive care unit.)

METHODS

An outbreak investigation, including detailed epidemiologic, clinical, and laboratory investigation, was performed.

RESULTS

A single strain of A. calcoaceticus anitratus was responsible for infection in all seven patients. All patients had tracheostomies, were in respiratory failure, and were ventilator dependent. Patients ranged in age from 27 to 81 years. No common causative variable or explanatory findings were present except that the same peak flow meter (manual weaning criteria machine) was used to facilitate weaning all seven patients from mechanical ventilation. Culture of the mouthpiece isolated a A. calcoaceticus anitratus strain with the identical susceptibility pattern and biochemical profile as that from the infected patients.

CONCLUSION

A. calcoaceticus anitratus was transmitted by a peak flow meter nosocomially to seven patients receiving mechanical ventilation. Disposable mouthpieces were introduced to prevent cross-contamination. A 2% glutaraldehyde solution was used to disinfect the machine between uses. No further outbreaks of A. calcoaceticus anitratus pneumonia were identified during 3 years of follow-up.

摘要

背景

在一家社区教学医院的重症监护病房发现了7例醋酸钙不动杆菌无硝亚种感染病例(第7例位于紧邻近期从重症监护病房转出的感染患者的降级病房)。

方法

开展了一次暴发调查,包括详细的流行病学、临床和实验室调查。

结果

所有7例患者的感染均由单一菌株的醋酸钙不动杆菌无硝亚种引起。所有患者均行气管切开术,存在呼吸衰竭,且依赖呼吸机。患者年龄在27至81岁之间。除使用同一台峰值流量计(手动撤机标准设备)辅助所有7例患者脱离机械通气外,未发现共同的致病变量或解释性发现。对口件进行培养,分离出一株醋酸钙不动杆菌无硝亚种,其药敏模式和生化特征与感染患者的菌株相同。

结论

醋酸钙不动杆菌无硝亚种通过峰值流量计在医院内传播给7例接受机械通气的患者。引入了一次性口件以防止交叉污染。在使用期间,使用2%的戊二醛溶液对设备进行消毒。在3年的随访期间,未发现醋酸钙不动杆菌无硝亚种肺炎的进一步暴发。