The Sheehan total knee endoprosthesis has been widely used since 1971. It takes the form of a semi-constrained hinge with intramedullary stems cemented into the femur and tibia for fixation. In Tayside the prosthesis has been in use since 1980. The clinical impression was that the prosthesis was not performing well and formal assessment of surviving prostheses was therefore carried out using the Hospital for Special Surgery (HSS) scoring system. Thirty-seven patients were available for follow-up, 15.6% of whom had good results while 40% had poor results according to this assessment. At review, 31% of patients had undergone revision surgery or were awaiting such surgery. This compares poorly with reported results of surface joint replacements. In the light of these results the authors feel that the Sheehan knee replacement is now obsolete.
自1971年以来,希恩全膝关节假体已被广泛使用。它采用半限制铰链形式,带有髓内柄,通过骨水泥固定于股骨和胫骨以实现固定。在泰赛德地区,该假体自1980年起投入使用。临床印象是该假体表现不佳,因此使用特殊外科医院(HSS)评分系统对留存的假体进行了正式评估。37例患者可供随访,根据该评估,其中15.6%的患者效果良好,而40%的患者效果不佳。复查时,31%的患者已接受翻修手术或正在等待此类手术。这与报道的表面关节置换结果相比很差。鉴于这些结果,作者认为希恩膝关节置换术现已过时。