Suppr超能文献

采用EMIT d.a.u.可卡因代谢物免疫分析法直接测定血清中的苯甲酰芽子碱。

Direct determination of benzoylecgonine in serum by EMIT d.a.u. cocaine metabolite immunoassay.

作者信息

Poklis A, Jortani S, Edinboro L E, Saady J J

机构信息

Department of Pathology, Medical College of Virginia, Virginia Commonwealth University, Richmond 23298-0597.

出版信息

J Anal Toxicol. 1994 Nov-Dec;18(7):419-22. doi: 10.1093/jat/18.7.419.

Abstract

An EMIT d.a.u. immunoassay for urine testing was applied on the Syva ETS Plus analyzer for the detection of the cocaine metabolite, benzoylecgonine (BE), in human serum. Serum was analyzed without prior extraction, concentration, or matrix modification. Calibrators and serum controls were prepared from EMIT d.a.u. calibrators that were reconstituted and diluted with EMIT Tox serum calibrator. The assay cutoff concentration for BE was 50 ng/mL. The within-run and between-run precisions of the assay were both less than 5%. Analysis of 162 patient serums yielded 43 BE positive results. All EMIT positive serum BE results were confirmed by gas chromatography-mass spectrometry. All patients with positive BE serums also had BE positive urine samples. Serum bilirubin and triglycerides as high as 38 mg/dL and 319 mg/dL, respectively, did not interfere with the assay. Modification of the EMIT urine assay allowed for a simple, rapid, and reliable method for the detection of BE in serum.

摘要

一种用于尿液检测的EMIT d.a.u.免疫测定法应用于Syva ETS Plus分析仪,以检测人血清中的可卡因代谢物苯甲酰芽子碱(BE)。血清分析无需事先提取、浓缩或基质改性。校准品和血清对照品由用EMIT Tox血清校准品复溶和稀释的EMIT d.a.u.校准品制备。BE的测定截断浓度为50 ng/mL。该测定法的批内和批间精密度均小于5%。对162份患者血清的分析产生了43份BE阳性结果。所有EMIT阳性血清BE结果均通过气相色谱-质谱法确认。所有BE血清阳性的患者尿液样本也呈BE阳性。血清胆红素和甘油三酯分别高达38 mg/dL和319 mg/dL时,不干扰该测定法。对EMIT尿液测定法的改进产生了一种简单、快速且可靠的血清中BE检测方法。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验