Focused US system for MR imaging-guided tumor ablation.

作者信息

Cline H E, Hynynen K, Watkins R D, Adams W J, Schenck J F, Ettinger R H, Freund W R, Vetro J P, Jolesz F A

机构信息

GE Corporate Research and Development, Schenectady, NY 12301.

出版信息

Radiology. 1995 Mar;194(3):731-7. doi: 10.1148/radiology.194.3.7862971.

Abstract

PURPOSE

To measure the performance characteristics of a focused ultrasound (US) system for magnetic resonance (MR) imaging-guided tumor ablation.

MATERIALS AND METHODS

The authors constructed a focused US system for MR imaging-guided tumor ablation. The location of the heated region and thermal dose were monitored with temperature-sensitive MR images obtained in phantoms and rabbit skeletal muscle after application of each sonic pulse.

RESULTS

The region heated by the focused ultrasound beam was within 1 mm of that observed on temperature-sensitive fast gradient-echo MR images of in vivo rabbit skeletal muscle. Analysis of heat flow and the rate of coagulation necrosis provided an estimate of the size of the ablated region that was in agreement with experimental findings.

CONCLUSION

MR imaging provides target definition and control for thermal therapy in regions of variable perfusion or in tissues that are not well characterized.