Imaging in the far-red with electronic light microscopy: requirements and limitations.

作者信息

Cullander C

机构信息

Department of Pharmacy, University of California, San Francisco 94143.

出版信息

J Microsc. 1994 Dec;176(Pt 3):281-6. doi: 10.1111/j.1365-2818.1994.tb03526.x.

Abstract

The acquisition of simultaneous dual confocal images with red and far-red light has both advantages (e.g. lower autofluorescence) and limitations. An understanding of these requisites is necessary to acquire high-quality images and to avoid the misinterpretation of experimental data. The poor detection of far-red light mandates a high optical transfer efficiency for the system, thus the transmittance of the objective lens and its axial and lateral chromatic aberration in the far-red are important factors for consideration. This technical note is an attempt to 'demystify' the process of filter set design for confocal microscopy by discussing the considerations that went into the construction of a filter set for use with the reagents cyanine 3.18 (Cy3) and cyanine 5.18 (Cy5), and thus to encourage users to look beyond the multi-purpose designs available commercially. The 568-nm laser line exciting Cy3 is at its emission maximum, which limits the collectable Cy3 fluorescence. High-transmission optical filters with sharp band pass cutoffs are thus desirable for maximum light throughout. Light path mirror efficiency rapidly degrades above 700 nm, but the loss of this portion of the Cy5 emission spectrum is acceptable since the fluorophore is very bright, and these very long wavelengths are also likely to introduce aberration. While resolution is decreased with far-red light, there is also greater penetration and less scattering, and it is thus possible to obtain high-quality images from deeper within the specimen.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验