Abstract

Analysis of blood gases and blood pH yield important information in many situations of clinical emergencies. We report on a patient in whom pre-analytic errors in blood gas and blood pH measurements resulted in unnecessary further investigations. We therefore studied various pre-analytic sources of error in blood pH and blood gas analysis. Delay in sample processing for more than one hour resulted in an increase of pO2 and pCO2 and a decrease of pH. Excess sodium heparin solution as an anticoagulant (> or = 10% of total volume) led to a significant decrease of pH and pCO2 and to an increase of pO2. Air bubbles (10% of total volume) left in the syringe for 10 min significantly increased pO2. For accurate estimations of pO2, pCO2 and pH, it is necessary to keep the heparin solution below 10% of total volume, to expel all air bubbles from the syringe and to process the blood sample within one hour. Instructions to medical staff on handling blood samples for blood gas analysis should include these possible sources of errors.

摘要

在许多临床紧急情况下,对血气和血液pH值的分析可提供重要信息。我们报告了一名患者,其血气和血液pH值测量中的分析前误差导致了不必要的进一步检查。因此,我们研究了血液pH值和血气分析中各种分析前误差来源。样本处理延迟超过一小时会导致pO2和pCO2升高以及pH值降低。过量的肝素钠溶液作为抗凝剂(≥总体积的10%)会导致pH值和pCO2显著降低以及pO2升高。注射器中留存10分钟的气泡(占总体积的10%)会使pO2显著升高。为准确估计pO2、pCO2和pH值,有必要使肝素溶液低于总体积的10%,排出注射器中的所有气泡,并在一小时内处理血样。向医务人员提供的关于处理用于血气分析的血样的说明应包括这些可能的误差来源。