Bartels E
Abteilung für Neurologie, Städtisches Krankenhaus München-Bogenhausen.
Ultraschall Med. 1993 Dec;14(6):272-8. doi: 10.1055/s-2007-1005260.
The purpose of this report is to describe, based on the experience with 50 patients, transcranial colour-coded duplex ultrasonography and to evaluate its practical potential in the routine examination in comparison with conventional transcranial Doppler ultrasonography. Under visual control, the Doppler angle between the ultrasonic beam and the flow direction can be adjusted. The mean angle of incidence in the middle cerebral artery was 33 degrees +/- 15 degrees, in the anterior cerebral artery 35 degrees +/- 17 degrees, in the posterior cerebral artery 45 degrees +/- 18 degrees and in the basilar artery 15 degrees +/- 14 degrees. The measurements of blood flow velocity under consideration of the angle of incidence resulted in 15%, 18%, 30% and 3% respectively higher velocity values. Examples of pathologic findings demonstrate the practical potential of this noninvasive method.
本报告旨在基于对50例患者的经验描述经颅彩色编码双功超声检查,并与传统经颅多普勒超声检查相比评估其在常规检查中的实际潜力。在视觉控制下,可调整超声束与血流方向之间的多普勒角度。大脑中动脉的平均入射角为33度±15度,大脑前动脉为35度±17度,大脑后动脉为45度±18度,基底动脉为15度±14度。考虑入射角后血流速度的测量结果显示速度值分别高出15%、18%、30%和3%。病理结果示例证明了这种非侵入性方法的实际潜力。