Kapil R P, Lee W, Wong N Y, Saxton P L, Lam G N
Drug Metabolism and Pharmacokinetics Section, Stine-Haskell Research Center, DuPont Merck Pharmaceutical Company, Newark, DE 19714.
J Chromatogr B Biomed Appl. 1994 Mar 4;653(2):221-6. doi: 10.1016/0378-4347(93)e0416-n.
A selective high-performance liquid chromatographic (HPLC) assay for a sigma receptor antagonist, DuP 734 (I), in rat plasma has been developed. Compound I and internal standard, XC031 (I.S.), were first extracted from plasma into an ethyl acetate-toluene mixture (3:7, v/v) and then back-extracted into freshly prepared phosphoric acid (0.03 M). Separation of I and I.S. with no interference from endogenous substances was achieved on a reversed-phase octyl column and detection was by UV at 229 nm. The mobile phase consisted of acetonitrile-glacial acetic acid-triethylamine-0.05 M ammonium acetate (670:4:2:2000, v/v). Using 0.5 ml of rat plasma for extraction, the limit of quantitation was 43 ng/ml and the assay was linear from 43 to 8536 ng/ml. The intra- and inter-day coefficients of variation ranged from 0.7 to 3.0%, and from 1.4 to 14.5%, respectively, over the entire concentration range. The accuracy was within 16.1% of the spiked concentrations. I was stable in frozen plasma at -20 degrees C for at least 68 days.