颌骨追踪系统JAWS-3D的精度。

Precision of the jaw tracking system JAWS-3D.

作者信息

Airoldi R L, Gallo L M, Palla S

机构信息

Department of Masticatory Disorders and Complete Dentures, Dental School, University of Zurich, Switzerland.

出版信息

J Orofac Pain. 1994 Spring;8(2):155-64.

PMID:7920351
Abstract

The purpose of this study was to measure, under laboratory conditions, the noise level of the optoelectronic jaw-tracking system JAWS-3D, which records the movement of the whole mandible from extraorally placed landmarks. Also, an assessment was made of its accuracy in measuring positions, angles, and velocities. The results showed that the accuracy of JAWS-3D decreased when the distance between the recorded point and the extraoral landmarks increased. The path of a point near to the landmarks was estimated with an error of 0.11%, whereas the error increased to 1.33% when the point was far from the landmarks. The maximum error in angle computation was 0.7 degrees. The velocities calculated by JAWS-3D corresponded closely to the actual ones: mean error of 3 mm/s for velocities up to 80 mm/s. Thereafter, the error increased to reach 26 mm/s at a speed of 210 mm/s.

摘要

本研究的目的是在实验室条件下,测量光电下颌追踪系统JAWS - 3D的噪声水平,该系统从口外放置的标志点记录整个下颌骨的运动。此外,还对其在测量位置、角度和速度方面的准确性进行了评估。结果表明,当记录点与口外标志点之间的距离增加时,JAWS - 3D的准确性会降低。靠近标志点的点的路径估计误差为0.11%,而当该点远离标志点时,误差增加到1.33%。角度计算的最大误差为0.7度。JAWS - 3D计算的速度与实际速度密切对应:速度高达80 mm/s时,平均误差为3 mm/s。此后,在速度为210 mm/s时,误差增加到26 mm/s。