Matsuda H, Yagishita A, Nakatsuji H, Miyazawa M
Division of Radiology and Nuclear Medicine, National Center Hospital for Mental, Nervous, and Muscular Disorders, NCNP.
Kaku Igaku. 1994 Aug;31(8):991-4.
A special bed and a new software program were developed for noninvasive regional cerebral blood flow (rCBF) measurements by 99mTc hexamethylpropylene amine oxime (99mTc-HMPAO) using only three-head SPECT system, Siemens MULTI-SPECT3. A large field gamma camera with a minimum of 38 cm-diameter is required for simultaneous acquisition of time activity curves of brain and aortic arch in this noninvasive method. A special bed with short width was designed and inserted obliquely into the gantry of the SPECT system to acquire both time activity curves of brain and aortic arch with the use of diagonal length (51 cm) of a gamma camera with 31 x 41 cm width. Moreover a software program was developed for processing a series of the graphical analysis of time activity curves, Lassen's linearization correction of SPECT images, and drawing of a rCBF map. This noninvasive rCBF measurement using only MULTISPECT3 in a short period with high spatial resolution is quite useful for routine clinical studies.
开发了一种特殊的床和一个新的软件程序,用于仅使用三头SPECT系统西门子MULTI-SPECT3通过99mTc六甲基丙烯胺肟(99mTc-HMPAO)进行无创性局部脑血流(rCBF)测量。在这种无创方法中,需要一台最小直径为38 cm的大视野伽马相机来同时采集大脑和主动脉弓的时间-活性曲线。设计了一种宽度较短的特殊床,并将其倾斜插入SPECT系统的机架中,以便利用宽度为31×41 cm的伽马相机的对角线长度(51 cm)采集大脑和主动脉弓的时间-活性曲线。此外,还开发了一个软件程序,用于处理时间-活性曲线的一系列图形分析、SPECT图像的拉森线性化校正以及rCBF图的绘制。这种仅使用MULTISPECT3在短时间内进行的具有高空间分辨率的无创rCBF测量对于常规临床研究非常有用。