Suppr超能文献

不同脊髓穿刺针进行硬膜穿刺所需的力及随后脑脊液漏出情况的比较。

Comparison of the force required for dural puncture with different spinal needles and subsequent leakage of cerebrospinal fluid.

作者信息

Westbrook J L, Uncles D R, Sitzman B T, Carrie L E

机构信息

Nuffield Department of Anesthesia, John Radcliffe Hospital, Headington, Oxford, United Kingdom.

出版信息

Anesth Analg. 1994 Oct;79(4):769-72. doi: 10.1213/00000539-199410000-00026.

Abstract

An in vitro model was used to determine the force required to pierce bovine dura with a range of new spinal needles and to measure the subsequent leakage rate of cerebrospinal fluid (CSF). A significantly greater force was required to pierce the dura with pencil-point style needles compared to Quincke needles of the same size. Quincke needles caused a greater loss of CSF than their pencil-point equivalents. The results suggest that there is not likely to be a significant reduction in postdural puncture headache (PDPH) using a 27-gauge pencil-point needle compared to a 25-gauge needle that may be easier to use. Different makes of the same design and gauge of needle showed significant differences in the amount of CSF leakage, which may influence the clinician's choice of needle.

摘要

采用体外模型来确定使用一系列新型脊柱穿刺针穿刺牛硬脑膜所需的力,并测量随后脑脊液(CSF)的漏出率。与相同尺寸的Quincke针相比,使用铅笔尖式针穿刺硬脑膜需要显著更大的力。Quincke针导致的脑脊液损失比同等铅笔尖式针更大。结果表明,与可能更容易使用的25号针相比,使用27号铅笔尖式针不太可能显著降低硬膜穿刺后头痛(PDPH)的发生率。相同设计和规格的不同品牌针在脑脊液漏出量上显示出显著差异,这可能会影响临床医生对针的选择。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验