In vitro modeling of spinal anesthesia. A digital video image processing technique and its application to catheter characterization.

作者信息

Robinson R A, Stewart S F, Myers M R, Lien L F, Rinaldi J R, Swisher J L, Drasner K

机构信息

Food and Drug Administration, Rockville, Maryland 20852.

出版信息

Anesthesiology. 1994 Oct;81(4):1053-60.

PMID:7943816
Abstract

BACKGROUND

Maldistribution of intrathecal local anesthetic has recently been implicated as a contributor to neurotoxic injury. In vitro modeling can be used to understand the distribution of anesthetic agents within the subarachnoid space. We describe an in vitro modeling technique that uses digital video image processing and its application to catheter injection of local anesthetic.

METHODS

A clear plastic model of the subarachnoid space, including a simulated spinal cord and cauda equina, was filled with lactated Ringer's solution. Phthalocyanine blue dye of known concentration was injected into the model through small-bore (28-G) and large-bore (18-G) catheters. Injections were performed at a variety of controlled rates and sacral catheter positions, and the propagation of dye throughout the model was recorded on videotape, digitized by computer, and converted to a two-dimensional image of dye concentration. A subset of data was compared with results obtained from spectrophotometric analysis.

RESULTS

There was a strong correlation (r = 0.98) between data obtained with analysis by digital video image processing and those obtained spectrophotometrically. Catheter size, catheter angle, and injection rate significantly influenced the distribution and peak concentration of simulated anesthetic. No major differences in distribution or peak concentration were observed with the two types of 28-G catheters.

CONCLUSIONS

The digital video image processing technique can be used to quantify anesthetic distribution rapidly within a model of the subarachnoid space without disturbing the distribution. The current results demonstrate a strong dependence of anesthetic distribution on catheter angle, catheter size, and injection rate. Comparisons between 28-G catheters suggest that the difference in reported incidence of cauda equina syndrome associated with different 28-G catheters cannot be explained on the basis of differences in anesthetic distribution.