Radiochemical purity testing for 99Tcm-exametazime: a comparison study for three-paper chromatography.

作者信息

Hung J C, Taggart T R, Wilson M E, Owens T P

机构信息

Department of Diagnostic Radiology, Mayo Clinic, Rochester, MN 55905.

出版信息

Nucl Med Commun. 1994 Jul;15(7):569-74. doi: 10.1097/00006231-199407000-00014.

Abstract

The standard radiochemical purity (RCP) determination uses a three-paper chromatography strip and solvent system (TP). The involvement of multiple strips for calculation of percentage primary, lipophilic 99Tcm-exametazime complex is tedious and time-consuming. The significant streaking of radioactivity on the ITLC/SG MEK strip of the TP method indicates that ITLC/SG MEK may not be an ideal system for RCP analysis of 99Tcm-exametazime. This study was undertaken to compare the standard TP method with two other proposed single-strip chromatography systems: Whatman 17 Chr paper with ethyl acetate (WE) and Gelman Solvent Saturation Pads paper with ether (GE). Our results showed that the solvent developing times (n = 55) and Rf values for TP, WE and GE were 130.4 +/- 9.0 s/0.5-1.0, 205.9 +/- 13.0 s/0.2-1.0 and 90.2 +/- 7.5 s/0.8-1.0, respectively. For RCP values ranging from 45.0 to 94.6% (n = 61), both WE and GE closely correlated with TP (r = 0.97 and 0.96). However, in the intermediate RCP range (i.e. 75-85%, n = 25), the false RCP acceptance rate (i.e. RCP > or = 80%) was 40% (10/25) for the WE method versus 4% (1/25) for the GE method. The GE method has the most clear separation of lipophilic 99Tcm-exametazime from other radiochemical impurities and offers the quickest RCP analysis for 99Tcm-exametazime with relatively accurate results.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验