Dekutoski M B, Schendel M J, Ogilvie J W, Olsewski J M, Wallace L J, Lewis J L
Twin Cities Scoliosis Spine Center, Department of Orthopaedic Surgery, University of Minnesota, Minneapolis.
Spine (Phila Pa 1976). 1994 Aug 1;19(15):1745-51. doi: 10.1097/00007632-199408000-00015.
For in vitro studies, there is no basis for choosing a "load control study" over a "displacement control" study. This study qualitatively compared results from in vitro and in vivo tests, allowing the authors to address the experimental assumptions that in vitro testing contributes to the understanding of the in vivo condition.
To compare motion changes at segments adjacent to fusions for in vitro and in vivo tests.
Investigators have measured the effects of spinal fusions on the adjacent segment in a human cadaver model and found greater adjacent facet joint load after fusions. Others have found significant increases in motion and facet loads at segments adjacent to in vitro lumbosacral and long fusions, when the same range of motion was repeated before and after immobilization of lumbar segments.
L2-L3 motion was measured in vitro by an instrumented spatial linkage under load and displacement control before and after immobilization of segments (L3-L7). In vivo, L2-L3 motion was measured while animals walked on a treadmill. L3-L7 was fused and the L2-L3 motion testing was repeated. The change in in vivo adjacent segment motion was qualitatively compared with the in vitro change under "load" and "displacement" control.
Under "load" control, in vitro facet motion did not significantly change after immobilization, whereas under "displacement" control, the facet motion significantly increased from 2.2 +/- 0.4 mm to 4.1 +/- 0.6 mm. Post-instrumentation, in vivo L2-L3 facet motion increased significantly. This change in vivo related better to the changes seen in the in vitro "displacement" control test than to the in vitro "load" control test.
对于体外研究,没有理由在“负荷控制研究”和“位移控制”研究之间进行选择。本研究定性地比较了体外和体内试验的结果,使作者能够探讨体外试验有助于理解体内情况的实验假设。
比较体外和体内试验中融合相邻节段的运动变化。
研究人员在人体尸体模型中测量了脊柱融合对相邻节段的影响,发现融合后相邻小关节负荷更大。其他人发现,在体外腰骶部和长节段融合的相邻节段,当在腰椎节段固定前后重复相同的运动范围时,运动和小关节负荷显著增加。
在节段(L3-L7)固定前后,通过在负荷和位移控制下的仪器化空间连杆在体外测量L2-L3运动。在体内,当动物在跑步机上行走时测量L2-L3运动。融合L3-L7并重复L2-L3运动测试。将体内相邻节段运动的变化与“负荷”和“位移”控制下的体外变化进行定性比较。
在“负荷”控制下,固定后体外小关节运动无显著变化,而在“位移”控制下,小关节运动从2.2±0.4mm显著增加到4.1±0.6mm。植入仪器后,体内L2-L3小关节运动显著增加。体内的这种变化与体外“位移”控制试验中观察到的变化比与体外“负荷”控制试验中观察到的变化更相关。