Ex vivo evaluation of ferromagnetism, heating, and artifacts produced by heart valve prostheses exposed to a 1.5-T MR system.

作者信息

Shellock F G, Morisoli S M

机构信息

Tower Imaging, Los Angeles, CA 90048.

出版信息

J Magn Reson Imaging. 1994 Sep-Oct;4(5):756-8. doi: 10.1002/jmri.1880040521.

Abstract

Ex vivo testing techniques were used to determine the ferromagnetic qualities of, presence of heating in, and artifacts produced by 13 different heart valve prostheses exposed to a 1.5-T (64-MHz) magnetic resonance (MR) system. None of the heart valve prostheses showed a measurable deflection in the 1.5-T static magnetic field. Only minimal artifacts were produced during MR imaging with a fast spoiled GRASS (gradient-recalled acquisition in the steady state) pulse sequence. The largest temperature changes measured during a "worst case" MR imaging sequence (estimated average specific absorption rate, 2.5 W/kg; estimated spatial peak specific absorption rate, 7.6 W/kg) were +0.2 degree C with the implant imaged "in air" and +0.3 degrees C with the implant imaged in normal saline. Therefore, MR procedures performed with a 1.5-T (64-MHz) MR system may be performed safely in patients with any of the 13 different heart valve prostheses evaluated in this study.