Frayne R, Holdsworth D W, Smith R F, Kasrai R, Larsen J P, Rutt B K
Department of Medical Biophysics, University of Western Ontario, London, Canada.
Magn Reson Med. 1994 Sep;32(3):410-7. doi: 10.1002/mrm.1910320318.
A turbine flow sensor for MR flow experiments has been evaluated using reference volume-flow rate measurements obtained using an electromagnetic (EM) flow meter measurements and simultaneous phase contrast (PC) MR acquisitions. After calibration, the device was found to have accuracy (compared with the EM flow meter), linearity, and precision of better than +/- 1%, +/- 3.5%, 3.5%, respectively, in constant flow mode (0 to 30 ml s-1). The frequency response of the flow sensor was flat (within +/- 10%) up to 13.9 Hz. Volume-flow rate measurements on constant and simulated physiologic flow waveforms were in close agreement with both the electromagnetic (EM) flow meter and the gated MR PC estimates.