Physical evaluation of computed radiography as a mammographic X-ray imaging system.

作者信息

Workman A, Cowen A R, Brettle D S

机构信息

FAXIL, University of Leeds, Academic Unit of Medical Physics, UK.

出版信息

Br J Radiol. 1994 Oct;67(802):988-96. doi: 10.1259/0007-1285-67-802-988.

Abstract

The physical imaging properties of a computed radiography (CR) system operating under mammographic exposure conditions have been measured. These measurements include modulation transfer function (MTF), sensitometric response and noise power spectrum (NPS). These figures were used to derive signal-to-noise ratio (SNR) descriptors of the system performance. The same measures were derived for a mammographic film-screen system. The SNR properties of the CR system indicate superior low frequency performance over a wider dynamic range than the film-screen system. The spatial frequency dependent SNR properties of the CR system are, however, limited. The results of a psychophysical test support the results obtained from physical measurements. These results indicate that despite its limited sharpness, details of dimensions relevant to diagnosis in mammography (circular details 0.25 mm diameter to represent microcalcifications and 3 mm diameter to represent small masses) are reproduced using CR with SNRs which are at least comparable with those of a mammographic film-screen system.