小肠灌肠造影检查对患者的辐射剂量。

Patient radiation doses from enteroclysis examinations.

作者信息

Hart D, Haggett P J, Boardman P, Nolan D J, Wall B F

机构信息

National Radiological Protection Board, Chilton, Didcot, Oxon, UK.

出版信息

Br J Radiol. 1994 Oct;67(802):997-1000. doi: 10.1259/0007-1285-67-802-997.

Abstract

Data relating to patient dose have been acquired for enteroclysis examinations (small bowel enemas) performed at the John Radcliffe Hospital, Oxford, on 23 adult patients. Dose-area products, fluoroscopy times and the number of radiographs taken are used to compare the examination procedure at the John Radcliffe Hospital with enteroclysis and barium follow-throughs performed elsewhere. The mean dose-area product for the 23 examinations was 6.8 Gy cm2 and the mean effective dose was estimated to be 1.5 mSv. These doses are intermediate between those arising from barium meals and barium enemas performed in the same room.

摘要

已获取牛津约翰拉德克利夫医院对23名成年患者进行小肠灌肠检查(小肠造影)时的患者剂量相关数据。使用剂量面积乘积、透视时间和所拍摄的X光片数量,将约翰拉德克利夫医院的检查程序与其他地方进行的小肠造影和钡剂通过检查进行比较。23次检查的平均剂量面积乘积为6.8 Gy cm²,估计平均有效剂量为1.5 mSv。这些剂量介于在同一房间进行的钡餐和钡灌肠检查所产生的剂量之间。