Coronary artery dissection caused by exit of the guidewire through the distal perfusion sidehole of an auto-perfusion angioplasty balloon catheter.

作者信息

Werns S W, Bates E R

机构信息

Division of Cardiology, University Hospital, University of Michigan Medical Center, Ann Arbor 48109.

出版信息

Cathet Cardiovasc Diagn. 1994 Sep;33(1):32-5. doi: 10.1002/ccd.1810330109.

Abstract

A previously unrecognized cause of coronary artery dissection is reported. A 67-year-old woman underwent angioplasty of the right coronary artery using an autoperfusion balloon catheter. Dissection occurred because the balloon catheter was advanced while the guidewire exited from one of the distal perfusion sideholes.

摘要

相似文献

8
[Catheter-induced dissection of the right coronary ostium and aortic root].
Z Kardiol. 1999 Jul;88(7):526-8. doi: 10.1007/s003920050318.
10
Surgical management of instrumentation-induced coronary artery dissection.
J Card Surg. 1995 Nov;10(6):626-31. doi: 10.1111/j.1540-8191.1995.tb00652.x.