[Digital storage phosphor radiography. Doses and image quality].

作者信息

Salvini E, Pedroli G, Montanari G, Pastori R, Crespi A, Zincone G

机构信息

Servizio di Radiologia e di Fisica Sanitaria, Ospedale San Gerardo, Monza, Milano.

出版信息

Radiol Med. 1994 Jun;87(6):847-51.

PMID:8041940
Abstract

Digital radiography (DR) based on storage phosphor technology is progressively replacing conventional screen-film system radiographic techniques (CR). However, many questions about image quality and dose reduction are still open. Thus, since DR spatial resolution is always lower than that of high quality screen-film images and image noise is conversely higher, for general radiography examinations especially, the signal-to-noise ratio is always lower with DR than with CR at the same dose level. However, the wide dynamic range, the linear response of storage phosphor detectors and the automatic read-out control of the digital system, enable lower DR X-ray imaging efficiency to be overcome, producing a nearly perfect readout of the image data every time an exposure is made. Moreover, post-processing can make the detection of abnormalities easier. This is why in many ROC clinical accuracy studies DR performed equally or better than CR. Good quality DR images, similar to conventional ones at the same dose levels, are therefore achievable in chest, abdomen, bone and soft-tissues examinations. Dose evaluation protocols are strongly needed in DR since overexposures due to technical mistakes may be overlooked; moreover, doses can be adjusted to the peculiar diagnostic need. The S values shown on any hard-copy image can be used to this purpose, even though it must be pointed out that they do not express the actual mean dose to the detector.

摘要

基于存储磷光体技术的数字射线摄影(DR)正在逐步取代传统的屏-片系统射线摄影技术(CR)。然而,关于图像质量和剂量降低的许多问题仍然没有解决。因此,由于DR的空间分辨率总是低于高质量屏-片图像,而图像噪声则相反更高,特别是对于一般的射线摄影检查,在相同剂量水平下,DR的信噪比总是低于CR。然而,存储磷光体探测器的宽动态范围、线性响应以及数字系统的自动读出控制,使得能够克服DR较低的X射线成像效率,每次曝光时都能几乎完美地读出图像数据。此外,后处理可以使异常检测更容易。这就是为什么在许多ROC临床准确性研究中DR的表现与CR相当或更好。因此,在胸部、腹部、骨骼和软组织检查中,可以获得与相同剂量水平下的传统图像相似的高质量DR图像。由于技术失误导致的过度曝光可能会被忽视,所以DR非常需要剂量评估方案;此外,剂量可以根据特殊的诊断需求进行调整。任何硬拷贝图像上显示的S值都可以用于此目的,尽管必须指出它们并不表示探测器实际接受的平均剂量。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验