Broudiscou L, Papon Y
INRA-Theix, station de recherche sur la nutrition des herbivores, Saint-Genès-Champanelle, France.
Reprod Nutr Dev. 1994;34(3):193-200.
The use of an electrode for the determination of ammonia in rumen and fermentor fluid samples has been evaluated. The amount of interfering volatile bases was negligible in our samples. The memory effect was only significant at the limits of the ammonia concentration range. The results from the electrode method were linearly correlated with the values from the standard colorimetric method (slope = 1.009; R = 0.998).
已对使用电极测定瘤胃液和发酵罐液样本中的氨进行了评估。在我们的样本中,干扰性挥发性碱的量可忽略不计。记忆效应仅在氨浓度范围的极限处显著。电极法的结果与标准比色法的值呈线性相关(斜率 = 1.009;R = 0.998)。