Theory of the oscillometric maximum and the systolic and diastolic detection ratios.

作者信息

Drzewiecki G, Hood R, Apple H

机构信息

Department of Biomedical Engineering, Rutgers, State University of New Jersey, Piscataway 08855-0909.

出版信息

Ann Biomed Eng. 1994 Jan-Feb;22(1):88-96. doi: 10.1007/BF02368225.

Abstract

It is proposed that the maximum in cuff pressure oscillations during oscillometry is due to the buckling of the brachial artery under a cuff. This theory is investigated by means of a mathematical model of oscillatometry that includes the mechanics of the occlusive arm cuff, the arterial pressure pulse waveform, and the mechanics of the brachial artery. A numerical solution is provided for the oscillations in cuff pressure for one cycle of cuff inflation and deflation. The buckling pressure is determined from actual arterial data and the von Mises buckling criteria. The buckling of an artery under a cuff occurs near -2 to 0 mm Hg transmural pressure. This effect corresponds with a maximum arterial compliance and maximum cuff pressure oscillations when cuff pressure is nearly equal to mean arterial pressure (MAP), in support of the suggested theory. The model was also found to demonstrate the basic characteristics of experimental oscillometry, such as an increasing and decreasing amplitude in oscillations as cuff pressure decreases, the oscillations that occur when cuff pressure is above systolic pressure, maximum oscillation amplitudes in the range of 1 to 4 mm Hg, and an oscillatory maximum at cuff pressure equal to MAP. These findings support the case that the model is representative of oscillometry. Finally, the model predicted values for the systolic and diastolic detection ratios of 0.593 and 0.717, respectively, similar to those found empirically. These ratios alter with blood pressure, but the tightness of the cuff wrap did not change their value.

摘要

有人提出,示波法测量过程中袖带压力振荡的最大值是由于袖带下方肱动脉的屈曲所致。通过一个示波法数学模型对该理论进行了研究,该模型包括阻塞性手臂袖带的力学、动脉压力脉搏波形以及肱动脉的力学。针对袖带充气和放气一个周期内的袖带压力振荡提供了数值解。屈曲压力根据实际动脉数据和冯·米塞斯屈曲准则确定。袖带下方动脉的屈曲发生在跨壁压力接近 -2 至 0 mmHg 时。当袖带压力几乎等于平均动脉压(MAP)时,这种效应与最大动脉顺应性和最大袖带压力振荡相对应,支持了所提出的理论。还发现该模型展示了实验示波法的基本特征,例如随着袖带压力降低振荡幅度先增大后减小、袖带压力高于收缩压时出现的振荡、1 至 4 mmHg 范围内的最大振荡幅度以及袖带压力等于 MAP 时的振荡最大值。这些发现支持了该模型代表示波法的观点。最后,该模型预测的收缩压和舒张压检测率分别为 0.593 和 0.717,与经验发现的值相似。这些比率会随血压变化,但袖带缠绕的松紧度不会改变其值。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验