A mean blood velocity statistic for the Doppler signal from a narrow ultrasound beam.

作者信息

Willink R, Evans D H

机构信息

Division of Medical Physics, Faculty of Medicine, Leicester Royal Infirmary, University of Leicester, England.

出版信息

IEEE Trans Biomed Eng. 1994 Apr;41(4):322-31. doi: 10.1109/10.284960.

Abstract

An easily calculable statistic proportional to the instantaneous spatial mean blood velocity through a vessel cross section is derived from Doppler power spectral estimates for the case where the Doppler beam is assumed to be of negligible thickness compared to the vessel diameter. This is an alternative statistic to that derived where uniform insonation is assumed, an assumption thought to be poorer in many real cases. The main requirement is that the velocity profile is monotonic increasing from the vessel wall to the vessel center. Errors in each statistic are compared for a variety of true beam dimensions, using a variety of velocity profiles, and the new statistic is shown to incur less error for Gaussian beam response profiles with a standard deviation less than 0.4 of the vessel radius, or for rectangular response profiles with a width less than 0.65 of the vessel diameter. If an estimate can be made of the true beam dimensions and vessel diameter, a weighted sum of the two statistics can give a more accurate estimate of mean velocity. The case of a beam displaced from the center of the vessel is also considered, and errors are found to be less than 4% for a displacement of 20% of the vessel radius.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验