Suppr超能文献

[有创血压测量。实际安全性]

[Invasive blood pressure measurements. Factual safety].

作者信息

Nielsen L H

机构信息

Anaestesiologisk intensiv afdeling, Aalborg Sygehus.

出版信息

Ugeskr Laeger. 1994 Aug 1;156(31):4450-3.

PMID:8066948
Abstract

Intra-arterial blood pressure measurement is often used in patients with unstable haemodynamics. The demand for accuracy in such measurements is high. Usually these demands are fulfilled, but situations can occur where the dynamic characteristics of the system are exceeded. In order to acknowledge this situation, one must be aware of these dynamic characteristics. The significance of the system's resonance frequency and damping is described. A method to control the usability of the system is described.

摘要

动脉内血压测量常用于血流动力学不稳定的患者。此类测量对准确性的要求很高。通常这些要求能够得到满足,但也可能出现系统动态特性被超越的情况。为了认识到这种情况,必须了解这些动态特性。文中描述了系统共振频率和阻尼的意义。还描述了一种控制该系统可用性的方法。