Suppr超能文献

Greenfield inferior vena cava filter dislodged during central venous catheter placement.

作者信息

Marelich G P, Tharratt R S

机构信息

Department of Internal Medicine, University of California, Davis Medical Center, Sacramento.

出版信息

Chest. 1994 Sep;106(3):957-9. doi: 10.1378/chest.106.3.957.

Abstract

Two cases of a Greenfield inferior vena cava (IVC) filter being dislodged during central venous catheter placement are described. This event appears to be related to J-tip guidewire use during venous catheterization. To our knowledge, this occurrence previously has not been reported. One of the two filters was dislodged from the IVC to a new position in the right atrium. Dislodging of the filter in these two particular cases did not give rise to clinical signs or symptoms.