Suppr超能文献

Metallosis caused by alumina ceramic screw fixation of grafted bone in a bipolar hip endoprosthesis.

作者信息

Watanabe M, Okumura H, Kihara Y, Shibata T

机构信息

Department of Orthopaedic Surgery, School of Medicine, Ehime University Shizukawa, Japan.

出版信息

Arch Orthop Trauma Surg. 1993;112(6):296-8. doi: 10.1007/BF00452969.

Abstract

We report a case of metallosis due to ceramic fragments in the inner bearing surface of the polyethylene cup of a bipolar hip prosthesis with fracture of the alumina ceramic screws used for grafted bone fixation. After bone-grafting acetabuloplasty with a Bateman bipolar endoprosthesis, gradual migration of the outer head occurred, and fracture of the screws was observed. Seven years after the operation, radiographs showed severe metallosis of the surrounding tissues. At revision, marked wear of the inner head was seen. Detailed examination revealed that many rough-edged ceramic particles originating from the fractured ceramic screws had become embedded in the inner surface of the polyethylene bearing insert. These extremely hard ceramic fragments had entered the prosthetic joint space during joint movement, thereafter causing continuous abrasion of the metal head. After revision, the concentration of the constituents of the prosthesis alloy in the serum decreased toward control levels. To prevent serious complications like those in the present case, screws should be inserted more than 1 cm from the outer head of a bipolar endoprosthesis, and the outer head should be removed as soon as possible if it comes into contact with the screws.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验