In this paper the measurement technique pulse oximetry is examined in 25 halothane-anaesthetized horses. Furthermore measures are presented which lead to a successful sensor placement at the tongue of the horse. The hemoglobin saturation determined by the pulse oximeter (SaO2) correlated very well with the hemoglobin saturation calculated by blood gas analysis (sO2). Nevertheless in the range of low saturation the pulse oximeter increasingly overestimates sO2. Pulse oximetry is an important progress in equine patient monitoring. A decline of oxygen saturation in the blood is detected immediately and the registration of the pulse amplitude renders a rough estimation of the quality of peripheral perfusion.
本文对25匹接受氟烷麻醉的马进行了脉搏血氧测定法测量技术的研究。此外,还介绍了一些措施,这些措施有助于在马的舌头成功放置传感器。脉搏血氧仪测定的血红蛋白饱和度(SaO2)与血气分析计算的血红蛋白饱和度(sO2)相关性非常好。然而,在低饱和度范围内,脉搏血氧仪对sO2的高估越来越明显。脉搏血氧测定法是马科动物患者监测方面的一项重要进展。能立即检测到血液中氧饱和度的下降,并且脉搏幅度的记录有助于对外周灌注质量进行粗略估计。