Clinical evaluation of a portable ultrasonic and a standard optical pachometer.

作者信息

Patel S, Stevenson R W

机构信息

Department of Vision Sciences, Glasgow Caledonian University, Scotland.

出版信息

Optom Vis Sci. 1994 Jan;71(1):43-6. doi: 10.1097/00006324-199401000-00009.

Abstract

The clinical performance of a portable ultrasonic pachometer was compared with a modified optical pachometer on a group of normal subjects. The mean central corneal thickness according to optical pachometry using the "edge-to-edge" method of optical doubling was 0.553 mm (SD = +/- 0.022). The mean central corneal thickness using the ultrasonic device was 0.506 mm (SD = +/- 0.039). The differences between the means was statistically significant (at 1% level of probability). Based on the light distribution within the corneal optical section viewed during optical pachometry a hypothesis is derived supporting the notion that optical pachometry tends to overestimate corneal thickness rather than ultrasonic pachometry underestimating corneal thickness.