Abstract

After dosimetric recording of the radium implants being in use at the radiological department of the gynecological clinics at Vienna, an attempt is made for the adaptation of cobalt-60 teletherapy to brachytherapy generally applied here and utilizing relatively high dosage. A biaxial, bisegmental small-angle rotation about two cranially convergent axes, combined with the radium doses, yields a relatively well equilibrated load on the parametrial tissue and on the essential areas of the lymph flow. The dose distribution is shown by transversal as well as by frontal sectional planes. If deviations of the position of the radium implants, always possible in an individual case, are verified by localization imaging, asymmetrical dose distributions can be calculated by a computer and compensated by means of different loads to the axes of rotation.

摘要

在对维也纳妇科诊所放射科正在使用的镭植入物进行剂量记录后,尝试使钴 - 60远距治疗适用于此处普遍应用且使用相对高剂量的近距离治疗。围绕两个向头侧汇聚的轴进行双轴、双节段小角度旋转,并结合镭剂量,可使子宫旁组织和淋巴引流的关键区域承受相对平衡的负荷。剂量分布通过横断面和额状断面显示。如果在个别情况下镭植入物的位置偏差总是可能出现的,通过定位成像加以验证,那么不对称的剂量分布可由计算机计算出来,并通过对旋转轴施加不同负荷来进行补偿。