关节造影术在评估有症状的非骨水泥型全髋关节假体松动方面的价值。

The value of contrast arthrography in assessing loosening of symptomatic uncemented total hip components.

作者信息

Barrack R L, Tanzer M, Kattapuram S V, Harris W H

机构信息

Orthopaedic Biomechanics Laboratory, Massachusetts General Hospital, Boston 02114.

出版信息

Skeletal Radiol. 1994 Jan;23(1):37-41. doi: 10.1007/BF00203699.

Abstract

Twenty-five patients with symptomatic uncemented total hip components were studied with contrast arthrography prior to surgical exploration. All but one had uncemented femoral stems and 16 had an uncemented acetabular component. As judged by the findings at surgery, on the femoral side the sensitivity, specificity, and accuracy of arthrography were 57%, 60%, and 58% respectively. There was a relatively high incidence of both false positives (17%) and false negatives (25%). On the acetabular side sensitivity, specificity, and accuracy were 29%, 89%, and 62.5%. False negatives were common (31%), while there was only one false positive. The results in this small series show that arthrography has distinct limitations in identifying the fixation status of uncemented total hip components.

摘要

对25例有症状的非骨水泥型全髋关节假体患者在手术探查前进行了关节造影研究。除1例患者外,其余患者均采用非骨水泥型股骨柄,16例采用非骨水泥型髋臼假体。根据手术结果判断,在股骨侧,关节造影的敏感性、特异性和准确性分别为57%、60%和58%。假阳性(17%)和假阴性(25%)的发生率相对较高。在髋臼侧,敏感性、特异性和准确性分别为29%、89%和62.5%。假阴性很常见(31%),而假阳性只有1例。这个小样本系列的结果表明,关节造影在确定非骨水泥型全髋关节假体的固定状态方面有明显的局限性。