Barrack R L, Tanzer M, Kattapuram S V, Harris W H
Orthopaedic Biomechanics Laboratory, Massachusetts General Hospital, Boston 02114.
Skeletal Radiol. 1994 Jan;23(1):37-41. doi: 10.1007/BF00203699.
Twenty-five patients with symptomatic uncemented total hip components were studied with contrast arthrography prior to surgical exploration. All but one had uncemented femoral stems and 16 had an uncemented acetabular component. As judged by the findings at surgery, on the femoral side the sensitivity, specificity, and accuracy of arthrography were 57%, 60%, and 58% respectively. There was a relatively high incidence of both false positives (17%) and false negatives (25%). On the acetabular side sensitivity, specificity, and accuracy were 29%, 89%, and 62.5%. False negatives were common (31%), while there was only one false positive. The results in this small series show that arthrography has distinct limitations in identifying the fixation status of uncemented total hip components.