Shapiro P, Attix F H, August L S, Theus R B, Rogers C C
Med Phys. 1976 Mar-Apr;3(2):87-90. doi: 10.1118/1.594208.
The displacement correction factor to be used for analysis of fast-neutron dosimetric measurements using air-filled EG and G tissue-equivalent ion chambers in a tissue-equivalent phantom has been investigated using the MANTA neutron radiotherapy beam generated by 35-MeV deuterons on a thick Be target. The displacement correction factor inferred from these measurements is 0.970 for the EG and G IC-17 (1.0-cm3) ion chamber, and is 0.989 for the EG and G IC-18 (0.1-cm3 ion chamber. This multiplicative displacement correction factor has no significant dependence on depth in the phantom or on neutron beam size.