[Artifactual defect of inferior myocardium on 123I-metaiodobenzylguanidine myocardial SPECT: characteristic findings and preventive method on phantom study].

作者信息

Kobayashi H, Terada S, Kanaya S, Kashikura K, Momose M, Hosoda S, Kusakabe K

机构信息

Department of Radiology, Tokyo Women's Medical College Hospital.

出版信息

Kaku Igaku. 1994 Apr;31(4):359-66.

PMID:8196232
Abstract

Discordance between planar and SPECT image of inferior myocardial defect on 123I-metaiodobenzylguanidine myocardial scintigraphy (MIBG) was occasionally observed in the clinical studies. The purpose of this study is to clarify the existence of artifactual myocardial defect on MIBG SPECT image. Phantom study was performed in various kinds of situation using triple gamma camera. Artifactual defect adjacent to the liver was significantly observed in the case of increased liver/heart uptake ratio more than 2:1. Artifactual defect was markedly observed in the 180 degrees than that in the 360 degrees SPECT acquisition. To avoid artifactual defect during SPECT acquisition, liver phantom was moved down to the foot direction (vanishing liver position). In this phantom position, artifactual defect adjacent to the liver was not observed on reconstructed myocardial SPECT image. The vanishing liver position was also applicable to human study. Human body was inclined to right side at an angle of 15 degrees, artifactual defect could be avoided. Further study is needed to analyze the cause of this artifactual defect and to conform the preventive method with vanishing liver position in the routine myocardial SPECT examination.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验