重症监护病房中患者因便携式胸部X光摄影而受到电离辐射。

Patient exposure to ionising radiation in the intensive care unit due to portable chest radiography.

作者信息

Lynch L, Bowen M, Malone L

机构信息

Department of Anaesthesia, Beaumont Hospital, Dublin.

出版信息

Ir J Med Sci. 1994 Mar;163(3):136-7. doi: 10.1007/BF02965973.

Abstract

The chest X-ray is the most commonly performed radiological examination in the intensive care unit. We used TLDs to measure the radiation exposure in 30 ICU patients due to portable chest radiography. The mean number of CXR was 3 (range 1-11). The mean surface entry dose at the xiphisternum was 1.8 mGy (range 0.43-5.14 mGy) per patient and 0.63 mGy per CXR. Very small amounts of radiation were detected at the symphysis pubis and in more than half of the patients no radiation was detected at this site. These values are well above accepted norms. Patient exposure may be reduced by ordering fewer X-rays or by changing to a faster screen-film combination.

摘要

胸部X光检查是重症监护病房中最常进行的放射学检查。我们使用热释光剂量计来测量30名重症监护病房患者因便携式胸部X光摄影所受到的辐射剂量。胸部X光检查的平均次数为3次(范围为1 - 11次)。每位患者剑突处的平均体表入射剂量为1.8毫戈瑞(范围为0.43 - 5.14毫戈瑞),每次胸部X光检查的平均体表入射剂量为0.63毫戈瑞。在耻骨联合处检测到的辐射量非常少,超过半数的患者在此部位未检测到辐射。这些数值远高于公认的标准。可以通过减少X光检查的开具次数或改用更快的增感屏 - 胶片组合来降低患者的辐射暴露。