Atkinson J B, Chamberlin K L, de Csepel J, Srikanth M
Division of Pediatric Surgery, Children's Hospital of Los Angeles, CA 90027.
J Pediatr Surg. 1994 Mar;29(3):379-80. doi: 10.1016/0022-3468(94)90571-1.
In pediatric patients, central line implantation via the subclavian vein can be complicated by the presence of a kink in the peel-away sheath. A technique is described, which has been used on 62 occasions to overcome this problem. Reinserting the catheter through the kinked peel-away sheath in concert with a guidewire that extends 3 to 4 cm beyond the catheter tip can ensure successful placement.