J Chromatogr B Biomed Appl. 1994 Mar 4;653(2):131-8. doi: 10.1016/0378-4347(93)e0430-x.
Several modifications of an HPLC-electrochemical assay method for plasma levels of norepinephrine (NE), epinephrine (EPI), dopamine (DA), dihydroxyphenylglycol (DHPG), dihydroxyphenylalanine (DOPA) and dihydroxyphenylacetic acid (DOPAC) that improve the accuracy and reliability of DHPG, DOPA, and DOPAC measurements are described. In batch alumina extractions, increasing the amount of alumina decreased analytical recoveries of DHPG, DOPA, and especially DOPAC, and increasing the strength of the eluting acid increased recoveries of these catechols, without affecting recoveries of the amines NE, EPI and DA. Refrigeration (4 degrees C) until injection stabilized DOPAC in aqueous solution and therefore improved the reproducibility of plasma DOPAC measurements. Circulation of chilled water (15 degrees C) around the column using a water jacket decreased variability in retention times of the catechols and thereby facilitated identification of peaks, while enhancing separation of DHPG from the solvent front. Use of 6-fluoro-DOPA and 6-fluoro-DOPAC as internal standards did not improve inter-assay reliability. We recommend that in assays of plasma catechols including DOPAC, small (5 mg), precisely measured amounts of alumina be used, with a relatively strong eluting solution (e.g. 0.04 M phosphoric acid-0.2 M acetic acid, 20:80, v/v), and that the samples be refrigerated until injection, with column temperature held constant at less than 20 degrees C.
本文描述了一种用于测定血浆中去甲肾上腺素(NE)、肾上腺素(EPI)、多巴胺(DA)、二羟基苯乙二醇(DHPG)、二羟基苯丙氨酸(DOPA)和二羟基苯乙酸(DOPAC)水平的高效液相色谱 - 电化学检测方法的几种改进,这些改进提高了DHPG、DOPA和DOPAC测量的准确性和可靠性。在批量氧化铝萃取中,增加氧化铝的量会降低DHPG、DOPA尤其是DOPAC的分析回收率,而增加洗脱酸的强度会提高这些儿茶酚的回收率,且不影响胺类NE、EPI和DA的回收率。注射前冷藏(4℃)可使水溶液中的DOPAC稳定,从而提高血浆DOPAC测量的重现性。使用水套使15℃的冷水在柱周围循环,可降低儿茶酚保留时间的变异性,从而便于峰的识别,同时增强DHPG与溶剂前沿的分离。使用6 - 氟 - DOPA和6 - 氟 - DOPAC作为内标并不能提高批间可靠性。我们建议在包括DOPAC在内的血浆儿茶酚测定中,使用少量(5mg)精确称量的氧化铝,以及相对较强的洗脱液(如0.04M磷酸 - 0.2M乙酸,20:80,v/v),并且样品在注射前冷藏,柱温保持在20℃以下恒定。