Lavery L A, Higgins K R, Ashry H R, Athanasiou K A
Podiatric Residency Training Program, University of Texas Health Science Center, San Antonio.
J Am Podiatr Med Assoc. 1993 Oct;83(10):557-62. doi: 10.7547/87507315-83-10-557.
The purpose of this study was to compare the structural characteristics of 2.0-mm polyglycolic acid pins and 2.0-mm Steinmann pins in oblique closing base wedge osteotomies of the first metatarsal bone commonly used to correct metatarsus primus varus. Six pairs of fresh frozen cadaveric first metatarsal bones were osteotomized, fixated with either absorbable or stainless steel 2.0-mm pins, cemented in a specimen-holding jig, and tested with the Bionix Material Testing System. There was no significant difference in the mean ultimate force, ultimate displacement, or structural stiffness when comparing 2.0-mm polyglycolic acid pins and Steinmann pins in this model. The structural characteristics of 2.0-mm absorbable pins were the same as stainless steel pins at the time of initial placement in the oblique closing base wedge osteotomy of the first metatarsal.
本研究的目的是比较2.0毫米聚乙醇酸针和2.0毫米斯氏针在常用于矫正第一跖骨内翻的第一跖骨斜行闭合基底楔形截骨术中的结构特性。对六对新鲜冷冻的尸体第一跖骨进行截骨,用可吸收或不锈钢2.0毫米针固定,粘结在标本固定夹具中,并用Bionix材料测试系统进行测试。在该模型中比较2.0毫米聚乙醇酸针和斯氏针时,平均极限力、极限位移或结构刚度没有显著差异。在第一跖骨斜行闭合基底楔形截骨术中最初置入时,2.0毫米可吸收针的结构特性与不锈钢针相同。