Suppr超能文献

用于测定血清中苯妥英和苯巴比妥的克隆酶供体免疫分析的评估。

Evaluation of the cloned enzyme donor immunoassay for measurement of phenytoin and phenobarbital in serum.

作者信息

van der Weide J, Luiting H J, Veefkind A H

机构信息

Department of Clinical Chemistry, Psychiatric Hospital Veldwijk, Ermelo, The Netherlands.

出版信息

Ther Drug Monit. 1993 Aug;15(4):344-8. doi: 10.1097/00007691-199308000-00014.

Abstract

We report the evaluation of the cloned enzyme donor immunoassay (CEDIA) for the estimation of phenytoin and phenobarbital in serum. The assays were performed with a Hitachi 911 analyzer. Intra-assay coefficients of variation were from 1.5 to 4.4% for phenytoin and 1.6 to 5.5% for phenobarbital. Interassay coefficients of variation ranged from 1.8 to 5.3% for phenytoin and 2.9 to 4.8% for phenobarbital. Linearity was satisfactory, with a recovery of 103% at 11 mg/L and 110% at 22 mg/L phenytoin and 93% at 14 mg/L and 101% at 40 mg/L for phenobarbital. The detection limit was 1.2 mg/L for phenytoin and 0.6 mg/L for phenobarbital. Results of this assay correlated well with those of a conventional high-performance liquid chromatography (HPLC) method; r = 0.99 for phenytoin (n = 47) and r = 0.98 for phenobarbital (n = 48). The CEDIA was easy to handle and especially suitable for short turn-around time application.

摘要

我们报告了用于估算血清中苯妥英和苯巴比妥的克隆酶供体免疫分析(CEDIA)的评估。这些分析使用日立911分析仪进行。苯妥英的批内变异系数为1.5%至4.4%,苯巴比妥的批内变异系数为1.6%至5.5%。苯妥英的批间变异系数为1.8%至5.3%,苯巴比妥的批间变异系数为2.9%至4.8%。线性良好,苯妥英在11 mg/L时回收率为103%,在22 mg/L时为110%;苯巴比妥在14 mg/L时回收率为93%,在40 mg/L时为101%。苯妥英的检测限为1.2 mg/L,苯巴比妥的检测限为0.6 mg/L。该分析结果与传统高效液相色谱(HPLC)方法的结果相关性良好;苯妥英的r = 0.99(n = 47),苯巴比妥的r = 0.98(n = 48)。CEDIA易于操作,特别适用于周转时间短的应用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验