Suppr超能文献

闭环手部抓握神经假体的自动调谐

Automated tuning of a closed-loop hand grasp neuroprosthesis.

作者信息

Lemay M A, Crago P E, Katorgi M, Chapman G J

机构信息

Department of Biomedical Engineering and Orthopaedics, Case Western Reserve University, Cleveland, OH 44106.

出版信息

IEEE Trans Biomed Eng. 1993 Jul;40(7):675-85. doi: 10.1109/10.237697.

Abstract

An automated tuning algorithm was developed to reduce the time and skill required to tune a closed-loop hand grasp neuroprosthesis. The time reduction results from simultaneous tuning of four gain parameters controlling the dynamic response of the system, and from automation of the calculation and decision processes. The new tuning method is therefore an automated parallel tuning method, replacing a manual sequential method in which only one parameter at a time was tuned. RMS error between the step input and the grasp output is minimized, with absence of oscillation as a constraint. The difference between the system's RMS ramp tracking errors for the two tuning methods was less than 1% of the ramp size regardless of the initial values of the parameters, implying that the tuning methods were equivalent. However, the parallel tuning method was faster and required fewer trials than the sequential method. The capability of the closed-loop system to regulate grasp output in the presence of disturbances was compared with the capability without feedback. Patients were instructed to either grasp an object at a certain force level or to match a certain grasp opening. They would then lock their command at a fixed value, and either remain immobile to test time dependence or pronate and supinate their forearm to test postural disturbances. With closed-loop control, the grasp output was better regulated in the presence of disturbances, with an average output variance 60% lower than without feedback control.

摘要

开发了一种自动调谐算法,以减少对闭环手部抓握神经假体进行调谐所需的时间和技能。时间的减少源于同时调整控制系统动态响应的四个增益参数,以及计算和决策过程的自动化。因此,新的调谐方法是一种自动并行调谐方法,取代了一次只调整一个参数的手动顺序方法。在无振荡约束的情况下,将阶跃输入与抓握输出之间的均方根误差降至最低。无论参数的初始值如何,两种调谐方法的系统均方根斜坡跟踪误差之间的差异小于斜坡大小的1%,这意味着两种调谐方法是等效的。然而,并行调谐方法比顺序方法更快,所需的试验次数更少。将闭环系统在存在干扰时调节抓握输出的能力与无反馈时的能力进行了比较。患者被指示以一定的力水平抓握物体或匹配一定的抓握开度。然后,他们将命令锁定在一个固定值,要么保持不动以测试时间依赖性,要么旋前和旋后前臂以测试姿势干扰。通过闭环控制,在存在干扰的情况下,抓握输出得到了更好的调节,平均输出方差比无反馈控制时低60%。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验