Tatcher M
Med Phys. 1976;3(5):351-2. doi: 10.1118/1.594265.
Scatter-air ratios are tabulated quantities which are based on measurement and which are used to calculate the dose to points in the body of a patient exposed to radiation fields of arbitrary shape. For computing purposes it is convenient to represent these quantities by an equation. We have found a nine-parameter empirical function which expresses scatter-air ratio data for 60Co gamma rays and for 4- and 6-MV x rays. To test the equation it was used to calculate tissue-air ratios for a range of rectangular fields. The results are in good agreement with published values.
散射线-空气比是基于测量得到的列表数据,用于计算暴露于任意形状辐射场的患者体内各点的剂量。为便于计算,用一个方程来表示这些数据很方便。我们已经找到了一个九参数经验函数,它能表达钴-60γ射线以及4兆伏和6兆伏X射线的散射线-空气比数据。为检验该方程,用它计算了一系列矩形野的组织-空气比。结果与已发表的值吻合良好。