Van Driessche W, De Smet P, Raskin G
Laboratory for Physiology, KU Leuven, Belgium.
Pflugers Arch. 1993 Oct;425(1-2):164-71. doi: 10.1007/BF00374517.
This paper describes an automatic method to measure cell height (h) of epithelia grown as monolayers on transparent filter supports. Tissues are mounted in an Ussing-type chamber enabling solution exchange on both sides. The apical and basal side of the epithelial cells are marked with fluorescent beads. The image of the fluospheres is captured with a video camera and processed by a computer-based video imaging system. One basal reference bead in a gelatin layer on the filter support and up to three beads attached at the apical surface are used to monitor changes in cell height of three cells simultaneously. The focusing of the microbeads is done automatically by moving the objective with a piezoelectric device mounted on the nosepiece of the microscope. The algorithm for locating the bead is based on the changes in fluorescent light intensity emitted by the fluospheres. The method has an accuracy higher than 0.1 micron and a time resolution as low as 6 s if measurements are restricted to one bead at the apical side. The method was tested on artificial model systems and used to measure volume changes in renal cultured epithelia (A6) after exposing the serosal surface to hypotonic solutions and replacing cell-impermeable sucrose by an organic compound (glycerol) with a smaller reflection coefficient. Serosal hypotonicity elicited a rapid volume increase followed by regulatory volume decrease, whereas the organic compound replacement caused a steady increase in cell volume.
本文描述了一种自动测量在透明滤膜支持物上形成单层生长的上皮细胞高度(h)的方法。将组织安装在Ussing型小室中,以便两侧进行溶液交换。上皮细胞的顶端和基底侧用荧光珠标记。用摄像机捕获荧光球的图像,并由基于计算机的视频成像系统进行处理。滤膜支持物上明胶层中的一个基底参考珠和附着在顶端表面的多达三个珠子用于同时监测三个细胞的细胞高度变化。通过使用安装在显微镜镜座上的压电装置移动物镜来自动完成微珠的聚焦。定位珠子的算法基于荧光球发出的荧光强度变化。如果测量仅限于顶端的一个珠子,该方法的精度高于0.1微米,时间分辨率低至6秒。该方法在人工模型系统上进行了测试,并用于测量将浆膜表面暴露于低渗溶液并用反射系数较小的有机化合物(甘油)替代细胞不可渗透的蔗糖后,肾培养上皮细胞(A6)的体积变化。浆膜低渗引起快速的体积增加,随后是调节性体积减少,而有机化合物替代导致细胞体积稳定增加。