Williamson J A, Webb R K, Cockings J, Morgan C
Department of Anaesthesia and Intensive Care, University of Adelaide, S.A.
Anaesth Intensive Care. 1993 Oct;21(5):551-7. doi: 10.1177/0310057X9302100510.
The first 2000 incidents reported to the Australian Incident Monitoring Study were analysed with respect to the role of the capnograph. One hundred and fifty-seven (8%) were first detected by a capnograph and there were a further 18 (1%) in which capnography was contributory. Of the 1256 incidents which occurred in association with general anaesthesia 48% were "human detected" and 52% "monitor detected". The capnograph was ranked second and detected 24% of these monitor detected incidents; this figure would have been nearly 30% if a correctly checked, calibrated capnograph had always been used. The capnograph is a "front-line" monitor for oesophageal intubation, failure of ventilation, anaesthetic circuit faults, gas embolism, sudden circulatory collapse and malignant hyperthermia. It is a valuable "back-up" monitor when other monitors (e.g. low pressure alarm, pulse oximeter) are not in use, are being used incorrectly or fail. Such situations, in order of frequency of detection were: circuit-leak, overpressure of the breathing circuit, bronchospasm, leak of ventilator-driving-gas into the patient circuit, aspiration and/or regurgitation and hypoventilation. There were 20 reports of "failure", over two-thirds of which would not have occurred with appropriate checking and calibration. Seven were due to gas sampling problems and 6 to apnoea alarm failure. Two circuit leaks and 2 faulty unidirectional valves were not detected; on 3 occasions problems occurred due to power failure, calibration problems, or misinterpretation of an alarm.(ABSTRACT TRUNCATED AT 250 WORDS)
针对二氧化碳监测仪的作用,对澳大利亚事件监测研究报告的前2000起事件进行了分析。其中157起(8%)最初由二氧化碳监测仪检测到,另有18起(1%)二氧化碳监测仪起到了辅助作用。在与全身麻醉相关的1256起事件中,48%为“人工检测到”,52%为“监测仪检测到”。二氧化碳监测仪排名第二,检测到了这些监测仪检测到的事件中的24%;如果一直使用经过正确检查和校准的二氧化碳监测仪,这一数字将接近30%。二氧化碳监测仪是食管插管、通气失败、麻醉回路故障、气体栓塞、突然循环衰竭和恶性高热的“一线”监测仪。当其他监测仪(如低压警报器、脉搏血氧仪)未使用、使用不当或出现故障时,它是一种有价值的“备用”监测仪。按检测频率排序,这些情况包括:回路泄漏、呼吸回路超压、支气管痉挛、呼吸机驱动气体泄漏到患者回路、误吸和/或反流以及通气不足。有20起“故障”报告,其中超过三分之二如果进行适当的检查和校准就不会发生。7起是由于气体采样问题,6起是由于呼吸暂停警报故障。2起回路泄漏和2个单向阀故障未被检测到;有3次问题是由于停电、校准问题或警报误判导致的。(摘要截断于250字)