Ciesielski B, Reinstein L E, Meek A G, Wielopolski L
Brookhaven National Laboratory, SUNY-SB, Radiation Therapy Facility, Upton, New York 11973.
Med Phys. 1993 Sep-Oct;20(5):1453-5. doi: 10.1118/1.597130.
Calculations of the energy response of an agar-alanine phantom dosimeter (AAPD) to electrons in the energy range of 0.15-20 MeV together with experimental results at 16 MeV are presented. It is shown that the sensitivity of the EPR dosimeter (EPR signal/Gray) is independent of alanine crystal size and varies less than 2%, in the electron energy range indicated. Thus, the measured free radical density distribution may be used directly as an indication of the absorbed dose distribution in an irradiated phantom.