Gary C K, Piestrup M A, Boyers D G, Pincus C I, Pantell R H, Rothbart G B
Adelphi Technology, Palo Alto, California 94306.
Med Phys. 1993 Sep-Oct;20(5):1527-35. doi: 10.1118/1.597116.
Channeling radiation could provide a viable source for digital energy subtraction angiography (DESA). A signal to noise ratio (SNR) of 6.2 for a resolution of 0.5 mm x 0.5 mm could be achieved using a 6-mA 100-ms 20-MeV electron-beam pulse and a diamond channeling crystal as the x-ray source. This article investigates the choice of a DESA contrast agent and the parameters of a channeling-radiation x-ray source to develop a channeling-radiation DESA imaging system. The production of dual-energy peaks, the maximum available x-ray flux, the advantages of an area exposure, the necessity of a mosaic Bragg-crystal filter to reduce patient dose, the optimal energy separation of the peaks for a quasi-monochromatic x-ray source, and the reduction of the signal from bone are discussed, leading to estimated SNRs and image resolution for a channeling-radiation imaging system. The computer analysis developed to calculate the image quality is also discussed.
沟道辐射可为数字能量减影血管造影(DESA)提供一种可行的源。使用6毫安、100毫秒、20兆电子伏的电子束脉冲和金刚石沟道晶体作为X射线源,对于0.5毫米×0.5毫米的分辨率可实现6.2的信噪比(SNR)。本文研究了DESA造影剂的选择以及沟道辐射X射线源的参数,以开发一种沟道辐射DESA成像系统。讨论了双能峰的产生、最大可用X射线通量、面积曝光的优势、使用镶嵌布拉格晶体滤波器降低患者剂量的必要性、准单色X射线源的峰的最佳能量分离以及骨信号的减少,从而得出沟道辐射成像系统的估计信噪比和图像分辨率。还讨论了为计算图像质量而开发的计算机分析。