Nasointestinal tube placement with a pH sensor feeding tube.

作者信息

Heiselman D E, Vidovich R R, Milkovich G, Black L D

机构信息

Akron General Medical Center, OH 44307.

出版信息

JPEN J Parenter Enteral Nutr. 1993 Nov-Dec;17(6):562-5. doi: 10.1177/0148607193017006562.

Abstract

Radiographic confirmation of enteral feeding tube placement is a common practice representing considerable expense and causing delay in the initiation of enteral nutrition therapy. We evaluated an enteral feeding tube with a pH sensor, which allows immediate verification of the location of the tube by assessment of the pH upon insertion. Insertion pHs were obtained for 24 intensive care unit patients requiring feeding tube placement. Placement was verified radiographically and compared with expected location on the basis of the pH profile. The radiograph and the insertion pH profile were in agreement in 87.5% (21 of 24) of the cases. Concomitant use of histamine blockers did not affect the ability of the pH sensor to detect placement accurately (Fisher's Exact Test, p 5.71) Use of these pH measurements eliminates the need for radiographic documentation of placement, provides a savings for the patient, and may be beneficial in promoting enteral feedings in critically ill patients.

摘要

通过X光检查确认肠内喂养管的放置位置是一种常见做法,这会产生相当高的费用,并导致肠内营养治疗开始时间延迟。我们评估了一种带有pH传感器的肠内喂养管,该传感器可在插入时通过评估pH值立即验证管子的位置。对24名需要放置喂养管的重症监护病房患者获取了插入时的pH值。通过X光检查验证放置位置,并根据pH值曲线与预期位置进行比较。在87.5%(24例中的21例)的病例中,X光片和插入时的pH值曲线一致。同时使用组胺阻滞剂并不影响pH传感器准确检测放置位置的能力(Fisher精确检验,p = 5.71)。使用这些pH测量值无需通过X光检查记录放置位置,为患者节省了费用,并且可能有助于促进重症患者的肠内喂养。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验